Affiliation:
1. Department of Chemistry, Howard L. Hunter Chemical Laboratories, Clemson University, Clemson, South Carolina 29634-1905
Abstract
A double quadrupole mass spectrometer has been assembled to perform collision-induced dissociation (CID) of molecular ions sampled from a radio-frequency (rf) powered glow discharge atomization/ionization source. Low-energy ( Elab = 20 and 40 eV) collisions with argon target atoms efficiently dissociate species of the form M2+, MAr+, MO+, and Ar2+, and residual gas species. Exemplary CID mass spectra illustrate the enhanced reduction of potential molecular isobars relative to scattering losses of analyte (atomic) ions.
Subject
Spectroscopy,Instrumentation
Cited by
55 articles.
订阅此论文施引文献