Simple, Inexpensive System for Controlled-Dispersion Flow Analysis in Inductively Coupled Plasma Spectrometry

Author:

Wu Min1,Ensman Robert1,Hieftje Gary M.1

Affiliation:

1. Department of Chemistry, Indiana University, Bloomington, Indiana 47405

Abstract

An inexpensive peristaltic pump has been modified for computer control and coupled to a commercial autosampler to provide completely automated sample introduction and calibration via controlled-dispersion flow injection. Precision of 4.2% RSD can be achieved with an injection volume of 10 μL or larger. With this device, controlled-dispersion flow analysis was used for computer-controlled standard calibration of Ca+2 from a single concentrated sample; a dilution of 180-fold of a 10-μg/mL Ca+2 sample was readily achieved. Intelligent control of the system was also realized through the implementation of a LabVIEW software system on a Macintosh computer. Advantages of the new system are the benefit of simplicity and low cost for flow-injection analysis, a reduction in sample consumption, and the convenience of performing single-standard calibration.

Publisher

SAGE Publications

Subject

Spectroscopy,Instrumentation

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Flow injection;Sample Introduction Systems in ICPMS and ICPOES;2020

2. Chapter 2 Flow injection techniques;Discrete Sample Introduction Techniques for Inductively Coupled Plasma Mass Spectrometry;2000

3. Recent developments in flow-injection atomic spectroscopy;Advances in Atomic Spectroscopy;1998

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3