Affiliation:
1. Departamento de Química, Unidad de Química Analítica, Universidad Autónoma de Barcelona, E-08193 Bellaterra, Barcelona, Spain
Abstract
Near-infrared diffuse reflectance spectroscopy (NIRS) with a fiber-optic probe was used for the determination of the active compound in a commercial pharmaceutical preparation. In order to reduce the strong scatter in the spectra and prevent scatter-induced changes in measurements from prevailing over concentration-induced changes, several data preprocessing methods were tested: normalization, derivatives, multiplicative scatter correction, standard normal variate, and detrending. The effectiveness for reducing the scattering of each data preprocessing was assessed, and the best results were obtained with the use of the second derivative. The effect of the treatments on the quantitation of the active compound by partial least-squares regression (PLSR) was studied, similar results being obtained in all cases, with a relative standard error of prediction lower than 1.55%.
Subject
Spectroscopy,Instrumentation
Cited by
67 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献