Affiliation:
1. Department of Chemistry, Indiana University, Bloomington, Indiana 47405
Abstract
A procedure is described for fabricating capillary orifice tips useful for the generation of uniform droplets. These generator tips operate at relatively low applied pressure, are relatively resistant to clogging, and incorporate orifice diameters reproducible to 1%. The method involves implanting a fused-silica capillary within a glass sleeve. When heated, the glass and fused silica wet to form a tight seal with no cracking caused by differences in thermal expansion. This technique permits the routine production of capillary tips with fixed orifice and outer diameters.
Subject
Spectroscopy,Instrumentation
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献