Affiliation:
1. Department of Chemistry, University of Alberta, Edmonton, Alberta, Canada T6G 2G2
Abstract
Details are presented for the construction of photodiode array (PDA) measurement systems from commercial components. The PDA systems described include the Hamamatsu S2304–1024Q, a Reticon 1024S using the RC1000 and RC1001 circuit boards, and a Reticon 1024S using the RC1024S circuit board. Detals are presented for computer-controlled clocking and timing circuits, ADC sub-systems, and Peltier cooling subsystems. The measurement characteristics (sensitivity and detection limits) for all arrays are intercompared with the use of analyte emission signals from an inductively coupled plasma.
Subject
Spectroscopy,Instrumentation
Cited by
14 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献