Affiliation:
1. Lawrence Livermore National Laboratory, Liuermore, California 94550
Abstract
The value of diffuse reflectance as an infrared or near-infrared spectroscopic sampling procedure has been limited by the low efficiency of accessories designed for it. In terms of signal-to-noise ratio, these average 2–6% for integrating spheres and 10–12% for various ellipsoidal mirror arrangements. Much better performances, up to 37% efficiency, can be obtained by optimizing a concentric confocal ellipsoidal mirror arrangement by using a very large central opening in the amular collector mirror, and adapting the throughput of the detector to the geometry of the collected beam.
Subject
Spectroscopy,Instrumentation
Reference4 articles.
1. Diffuse reflectance measurements by infrared Fourier transform spectrometry
2. Wendlandt W. M. and Hecht H. G., Reflectance Spectroscopy (Interscience, New York, 1966), pp. 253–274.
3. Wendlandt W. M. and Hecht H. G., Reflectance Spectroscopy (Interscience, New York, 1966), pp. 46–89.
Cited by
15 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献