Affiliation:
1. Department of Applied Chemistry, College of Engineering, University of Osaka Prefecture, Sakai, Osaka 591, Japan
Abstract
A sensitive method is described for the determination of trace amounts of tin, based on continuous sodium borohydride reduction of tin in solution to gaseous stannane. The stannane generated is introduced into a relatively low-powered (1.6 kW) inductively coupled argon plasma source. The effects of a range of acids and interelement effects are investigated. Mixtures of nitric and succinic acids and of nitric and malic acids are the most suitable reaction media, especially in the presence of interferents. The method of standard additions is recommended for accurate tin determination. Compared to a conventional solution nebulization, the present system gives a sensitivity increase of a factor of approximately 3 orders of magnitude. The detection limit for tin is 0.05 ng/ml and precision values at 20 and 50 ng/ml Sn are less than 3% relative standard deviation. The linear calibration range spans nearly 5 orders of magnitude. The proposed method is applied to the determination of tin in low alloyed steels and in National Bureau of Standards standard reference materials (Cu-Ni alloy, orchard leaves, wheat flour, and rice flour).
Subject
Spectroscopy,Instrumentation
Cited by
39 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献