Affiliation:
1. Department of Chemistry, The George Washington University, Washington, D.C. 20052
Abstract
A tandem helium plasma was formed by placing a water-cooled, grounded electrode above a helium inductively coupled plasma (He ICP). The tandem plasma could be operated at 250 to 700 W. Detection limits of F, Cl, Br, S, and C at 500 W were comparable or superior to those from a He ICP operated at 1500 W. At 500 W, the electron number density for the tandem helium plasma was greater than that for the He ICP at 1500 W, but similar excitation temperatures and roughly similar rotational temperatures were obtained.
Subject
Spectroscopy,Instrumentation
Cited by
23 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献