Affiliation:
1. Department of Chemistry, Indiana University, Bloomington, Indiana 47405-4001
Abstract
Emission spectra in the UV-visible and near-infrared (NIR) spectral regions are presented for several nonmetals introduced as aqueous solutions by a glass-frit nebulizer. Detectability improves dramatically for some nonmetals with the use of a desolvation system to reduce the amount of water vapor entering the plasma. Detection limits for P, I, S, Cl, Br, and C are in the low- to sub-ppm range. Good precision (RSD <5%) and a dynamic range of 2–3 decades were observed for I and P. Easily ionizable elements such as K, Na, and Li can enhance nonmetal emission by as much as 30%.
Subject
Spectroscopy,Instrumentation
Cited by
38 articles.