Affiliation:
1. Varian Techtron Pty. Limited, 679 Springvale Road, Mulgrave, Victoria, 3170, Australia
Abstract
The design and performance characteristics of a continuous-flow vapor generation accessory for atomic absorption spectrophotometry are reported. When used with a conventional atomic absorption spectrophotometer, the accessory provides analytical sensitivity for mercury, arsenic, selenium, antimony, bismuth, tellurium, and tin equal or superior to that of a batch-type vapor generation device. Sample presentation, sample consumption, precision of measurement, and simplicity of operation are similar to those of flame atomic absorption with a pneumatic nebulizer. Details of the construction of the continuous-flow manifold, gas/liquid separator, and absorption cells are discussed, and the effects of reagent concentrations and carrier gas flow rates are shown. Characteristic concentrations and analytical curves for the above elements are also presented.
Subject
Spectroscopy,Instrumentation
Cited by
55 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献