Affiliation:
1. Ames Laboratory-USDOE and Department of Chemistry, Iowa State University, Ames, Iowa 50011
Abstract
Laser-based photoacoustic densitometry was applied to two-dimensional analysis in thin-layer chromatography. An acousto-optic device was used to provide both intensity modulation and a rastering scan. The spatial resolution is 25 × 60 spots for an area of 25 × 50 mm on the thin-layer chromatographic plate. Three detection modes, normal modulation, resonant modulation, and unmodulated rapid scan modes were compared. Detection limits in the most sensitive mode and in the fastest mode are 350 pg with scan time of 153 s and 6.9 ng with scan time of 1.5 s, respectively.
Subject
Spectroscopy,Instrumentation
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献