Abstract
Fiber optics were used to interface a Raman spectrometer to a long (1 m) sample tube, with the objective being increased sensitivity. Internal reflection of the laser light and the Raman scatter within the sample tube permitted a long solution length to be sampled, increasing the Raman sensitivity by factors of 30–50 over conventional capillary tube sampling systems. In addition, the sample was subjected to much lower power densities than with systems employing a focused beam, thus minimizing radiation damage. Detection limits of 10−9 to 10−8 M were achieved for resonance Raman scatterers, and normal Raman scatterers could be detected at the 1 × 10−5 M level.
Subject
Spectroscopy,Instrumentation