Affiliation:
1. Laboratoire de Physicochimie Industrielle Bat. 401, INSA 20, Avenue A. Einstein, 69621 Villeurbanne Cedex, France
Abstract
Different high-frequency generators have been studied in the 5 to 56 MHz range in order to observe the influence of frequency on spectral characteristics of inductively coupled plasmas used in emission spectroscopy. Higher frequencies tended to yield decreased excitation temperatures and electron number densities and therefore decreased line and background continuum intensities. Signal-to-noise ratios were improved ranging from 1 to 8 MHz going from 27 to 56 MHz with similar improvements in detection limits. Precision, working curve linearity, chemical interferences, and the ability to atomize high salt concentration samples were unaffected by frequency.
Subject
Spectroscopy,Instrumentation
Cited by
49 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献