Affiliation:
1. Department of Chemistry, The George Washington University, Washington, D.C. 20052
Abstract
Two cyclone spray chambers of different sizes are evaluated for inductively coupled plasma/atomic emission spectrometry. On the basis of a simplified mathematical model, the cut-off diameters of the aerosol droplets are estimated to be 2.2 to 4.3 and 1.8 to 3.5 μm for the large and the small cyclones, respectively. The sample consumption rate can be reduced to about 50 μL/min when the cyclone chamber is used in a recycling mode. In general, the signal-to-background ratios, detection limits, and precisions of the analyte signal intensities obtained with the small cyclone chamber are slightly superior to those achieved with a Scott spray chamber and a recycling gravitational sedimentation chamber. The performance of the recycling cyclone chamber is also evaluated for the determination of copper in the NBS freeze-dried urine.
Subject
Spectroscopy,Instrumentation
Cited by
16 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献