Affiliation:
1. The Department of Chemistry, Willard Hall, Kansas State University, Manhattan, Kansas 66506
Abstract
The successful application of LC-SLM Hadamard transform spectrometry as a simultaneous multiwavelength detection system to Raman spectroscopy is presented. Multiplexed Raman data are obtained with the use of an Ar+ laser lasing at 514.5 nm and a room-temperature silicon photodiode. A conventional 90° scattering geometry is employed for measurements. It is demonstrated that the LC-SLM Hadamard transform Raman spectrometer possesses the capability of performing spectral subtraction and the ability to obtain depolarization ratios of Raman bands, and can function as a selectively tunable optical filter for both Rayleigh line rejection and optical band notching. It is also demonstrated that, for the Hadamard transform Raman experiment, the silicon photodiode used as the detector produces spectra with slightly better signal-to-noise ratios than those obtained with the photomultiplier tube (PMT) used as the detector, although the PMT shows an increase in sensitivity.
Subject
Spectroscopy,Instrumentation
Cited by
48 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献