Validity of Hand Held Fundus Camera by Optometrist using Slit lamp 90 d bio microscopy as a reference standard for screening of Diabetes Retinopathy

Author:

Ahsan Shahid,Haseeb Uzma,Memon Muhammad Saleh,Bukhari Sadia,Mahmood Tauseef,Fahim Muhammad Faisal

Abstract

Objective: To find out the validity of hand-held fundus camera by optometrist using slit lamp 90D biomicroscopy as reference standard for screening of diabetes retinopathy. Method: The observational cross-sectional study was conducted at the diabetic clinic of Al-Ibrahim Eye Hospital, Karachi, from August 2020 to May 2021, and comprised diabetics of either gender aged >16 years visiting the outpatient department. Un-dilated fundus photograph of both eyes were taken with non-mydriatic fundus camera. Pupils were then mid-dilated with one drop of tropicamide 1% before capturing retinal images by handheld fundus camera by another optometrist. Both the optometrists identified and recorded the presence and absence of diabetic retinopathy. Subsequently, a retinal specialist examined the fundus with slit lamp 90 D biomicroscopy. Data was analysed using SPSS 23. Results: Of the 500 subjects, 291(58.2%) were males and 209(41.8%) were females. The overall mean age was 54.49±9.16 years (range: 16-83 years). Of the 1000 eyes, fundus was not readable in 130(13%) by hand-held fundus camera, 296(29.6%) eyes by non-mydriatic fundus camera and 76(7.6%) eyes by slit lamp. Sensitivity and specificity of hand-held fundus camera compared to non-mydriatic fundus camera was 89.86% and 80.36%, respectively. In comparison with slit lamp, the sensitivity was 91.71% and specificity was 71.10%. Kappa statistic for diabetic retinopathy detection by hand-held fundus camera versus non-mydriatic fundus camera was 0.705, indicating substantial agreement. Kappa statistic for diabetic retinopathy detection with hand-held fundus camera versus slit lamp (standard) was 0.609, indicating good agreement. Conclusion: Handheld fundus camera with semi-dilated pupil was found to be a valid screening tool in the hand of optometrist for preliminary screening of diabetic retinopathy. Key Words: Retinal screening, Hand-held fundus camera, Optometrist, Sensitivity, Specificity.

Publisher

Pakistan Medical Association

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3