Author:
Wu Hong-Peng ,Dong Lei ,Zheng Hua-Dan ,Liu Yan-Yan ,Ma Wei-Guang ,Zhang Lei ,Wang Wu-Yi ,Zhu Qing-Ke ,Yin Wang-Bao ,Jia Suo-Tang , ,
Abstract
A trace gas sensor, based on quartz-enhanced photoacoustic spectroscopy (QEPAS) with two non-resonant micro-tubes, was designed to detect the ammonia concentration in impure helium. Unlike the traditional micro-resonator, the non-resonant micro-tubes are used to confine the sound wave, but do not exhibit a well-defined resonant behavior. Such a design makes the dimension of the spectrophone much smaller than the micro-resonant configuration, which facilitates the optical alignment. Signal and noise, that were dependent on gas pressure, were also investigated to optimize sensor performance. With the optimal sensor parameters and the optimal gas pressure, the detection sensitivity was found to be 463 ppb (1 , 1 s averaging time), which corresponds to the normalized absorption sensitivity of 4.310-9cm-1W/Hz.
Publisher
Acta Physica Sinica, Chinese Physical Society and Institute of Physics, Chinese Academy of Sciences
Subject
General Physics and Astronomy
Reference16 articles.
1. Li Z Y, Wang H H, Jiang N, Cheng S L, Zhao L, Yu X 2009 Acta Phys.Sin. 58 3821 (in Chinese) [李政颖, 王洪海, 姜宁, 程松林, 赵磊, 余鑫 2009 物理学报 58 3821]
2. Liu Z M, Liu W Q, Gao M G, Tong J J, Zhang T S, Xu L, Wei X L 2008 Chin. Phys. B 17 4184
3. Kosterev A A, Tittel F K, Serebryakov D V, Malinovsky A L, Morozov I V 2005 Rev. Sci. Instrum. 76 043105
4. Kosterev A A, Bakhirkin Y A, Curl R F, Tittel F K 2002 Opt. Lett. 27 1902
5. Dong L, Kosterev A A, Thomazy D, Tittel F K 2010 Appl. Phys. B 100 627
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献