Quantitative analysis of accuracy of concentration inversion under different temperature and pressure
-
Published:2021
Issue:7
Volume:70
Page:073201
-
ISSN:1000-3290
-
Container-title:Acta Physica Sinica
-
language:
-
Short-container-title:Acta Phys. Sin.
Author:
Wang Yu-Hao,Liu Jian-Guo,Xu Liang,Liu Wen-Qing,Song Qing-li,Jin Ling,Xu Han-Yang, , ,
Abstract
The CO<sub>2</sub>, CH<sub>4</sub> and other greenhouse gases are measured by using a home-made Fourier transform infrared spectrometer at the Longfengshan atmospheric background station. Compared with the measurement results of the instrument from the background station which meets the standards of the World Meteorological Organization, the correlation coefficient and the root mean square error of the CO<sub>2</sub> concentration value are 0.9576 and 18.6015, so the measurement results from the home-made instrument are reliable. In the home-made instrument the calibration spectrum of standard temperature and the calibration spectrum of stand pressure are used to invert the concentration. With the temperature changing, the temperature of the measured gas will vary, thus resulting in error. The research of environmental variable factors can improve the accuracy of concentration inversion. For example, compared with CO<sub>2</sub> absorption spectrum under 296 K, the CO<sub>2</sub> absorption spectrum under 297 K will have 1.8% spectrum deviation and its inversion concentration error is 0.41%. This is the main cause of inversion concentration error. Based on the above analysis, the absorption cross section is calculated by using the high-resolution transmission molecular absorption database parameters. Combining with the instrument line shape, the calibration spectra at different temperatures and pressures can be obtained. The calibration spectra at different temperatures and pressures are used to calibrate the concentration inversion. After calibration, compared with the measurement results of the background station instrument, the correlation coefficient and the root mean square error of the CO<sub>2</sub> concentration value are 0.9637 and 6.7803. The correlation coefficient of CO<sub>2</sub> concentration value measured by self-developed instrument is improved and root mean square error is reduced. The result shows that the calibration algorithm enhances the accuracy of the measurement results to a certain extent. The above results illustrate the reliability of the home-made FTIR instrument and this experiment provides important data, which lay the foundation forstudying the home-made Fourier transform infrared spectrometer. Of course, improvement can be made in the following areas. Other minor factors may affect the effect of the inversion algorithm. The concentration inversion will have subtle differences at different bands of calibration spectra. So in order to improve the measurement accuracy, we need to choice more reasonable band inversion and more precise parameters from the high-resolution transmission molecular absorption database.
Publisher
Acta Physica Sinica, Chinese Physical Society and Institute of Physics, Chinese Academy of Sciences
Subject
General Physics and Astronomy
Reference20 articles.
1. Keeling C D, Whorf T P, Wahlen M, Van der Plichtt J 1995 Nature 375 666 2. Hodgkinson J, Smith R, Ho W O, Saffell J R, Tatam R P 2013 Sens. Actuators, B 186 580 3. Chen H, Winderlich J, Gerbig C, Hoefer A, Rella C W, Crosson E R, Van Pelt A D, Steinbach J, Kolle O, Beck V, Daube B C, Gottlieb E W, Chow V Y, Santoni G W, Wofsy S C 2010 Atmos. Meas. Tech. 3 375 4. Cheng S Y, Xu L, Cao M G, Jin L, Li S, Feng S X, Liu J G, Liu W Q 2013 Acta Phys. Sin. 62 124206 程巳阳, 徐亮, 高闽光, 金岭, 李胜, 冯书香, 刘建国, 刘文清 2013 物理学报 62 124206 5. Zhang C, Liu C, Hu Q, Cai Z, Su W, Xia C, Zhu Y, Wang S, Liu J 2019 Light- Sci. Appl. 8 1
Cited by
2 articles.