Author:
Zhuang Yi-Qi ,Bao Jun-Lin ,Sun Peng ,Wang Ting-Lan ,Chen Wen-Hao ,Du Lei ,He Liang ,Chen Hua ,
Abstract
The limitations to shot noise measurement methods based on superconducting quantum interference device (SQUID) and superconductivity-insulation-superconductor (SIS) Josephson junction are pointed out, and a method to measure the shot noises of conventional electronic devices is proposed. Shot noise characteristics of conventional electronic devices are analyzed, and then a low-temperature measurement system is established. By using a double-shielding construction and low noise preamplifier, the test system can achieve a good electromagnetic interference shielding and low background noise. The theoretical and the experimental results of shot noises in diodes at 10 K are in good agreement with each other. The accuracy and the credibility of measurement system are proved.
Publisher
Acta Physica Sinica, Chinese Physical Society and Institute of Physics, Chinese Academy of Sciences
Subject
General Physics and Astronomy
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献