Author:
Wang Lin-Jun ,Liu Jian-Min ,Su Qing-Feng ,Shi Wei-Min ,Xia Yi-Ben ,
Abstract
The influences of the applied bias voltage and the pre-irradiation treatment by alpha particles on the electrical properties of sandwich structural diamond film detectors under 5.5 MeV 241Am alpha-particle irradiation were investigated. Results of current- voltage(I-V) and pulse height distribution measurements showed that the dark current of the diamond film detector would increase due to the pre-irradiation by alpha particles. Under the alpha irradiation, the detector under negative bias voltage had a higher response current and a better signal-to-noise ratio than that under a positive bias. Raman scattering studies directly demonstrated that the above phenomenon resulted mainly from the different structural imperfection distributions along the thickness direction. An energy resolution of about 25.0% was obtained for the detector under a negative bias voltage and 38.4% under a positive bias voltage. With increasing alpha-particle irradiation time, both the response current and the charge collection efficiency increased obviously.
Publisher
Acta Physica Sinica, Chinese Physical Society and Institute of Physics, Chinese Academy of Sciences
Subject
General Physics and Astronomy
Cited by
8 articles.