Impact on diabetes control and patient-reported outcomes of a newer implantable continuous glucose monitoring system (Eversense® CGM System): a single-centre retro- and prospective observational study

Author:

Dimitri Guy Rohner ,Lukas Burget ,Christoph Henzen ,Stefan Fischli

Abstract

AIMS OF THE STUDY: The Eversense® CGM System is the first and only continuous glucose monitoring system (CGMS) that uses a fully subcutaneous implanted sensor. This study aimed to evaluate effectiveness, safety and patient-reported outcomes in patients using the Eversense® CGM System in a realistic clinical setting, assessed at a single Swiss diabetes centre (Luzerner Kantonsspital) with prolonged follow-up. METHODS: This was a prospective and retrospective observational study that included patients with type 1 diabetes mellitus in whom at least one Eversense® glucose sensor was implanted between 2017 and 2022. The primary endpoint was the change in HbA1c levels from the baseline (before implantation of the sensor) to 6 ± 2 and 12 ± 2 months and the last follow-up (newest available value) after implantation. The secondary outcome measures were the number of premature sensor breakdowns, adverse events related to the implantation procedure (infection, bleeding, difficulties with implantation or explantation) and patient-related outcomes (assessed with a questionnaire). RESULTS: A total of 33 patients participated in this study. The median follow-up time was 50 (IQR 22.3–58.5) months. In total, 178 sensor implantations were performed. Valid HbA1c results were available for 26 participants. Compared to the baseline values, HbA1c levels at 6 and 12 months and the last follow-up changed by –0.25%, –0.45 and –0.2 (p = 0.278, 0.308 and 0.296, respectively). We recorded 16 (9%) premature sensor breakdowns, all occurring between 2019 and 2020. Apart from one late-onset infection and four complicated sensor removals, no major complications were assessed. The results of the questionnaire showed a subjective improvement in hypoglycaemia rates, a better perception of hypoglycaemia and the impression of better diabetes management. Common issues with the device reported by the patients were technical errors (connection problems) and problems with the removal procedure. CONCLUSIONS: The use of the Eversense® CGM System resulted in changes in HbA1c of between –0.2% and –0.45%. The rate of premature sensor breakdown was low. Major complications following sensor implantation or removal were absent, apart from one case of infection and four cases of complicated removal. Patient-reported outcomes with the Eversense® CGM System showed a subjective positive impact on hypoglycaemia rates, greater confidence in managing hypoglycaemia and diabetes in general, and easy handling of the transmitter and mobile app. Technical issues must be considered but are nowadays, with the use of the newest sensor generation, very rare.

Publisher

SMW Supporting Association

Reference19 articles.

1. Lachin JM, Genuth S, Cleary P, Davis MD, Nathan DM; Diabetes Control and Complications Trial/Epidemiology of Diabetes Interventions and Complications Research Group. Retinopathy and nephropathy in patients with type 1 diabetes four years after a trial of intensive therapy. N Engl J Med. 2000 Feb;342(6):381–9. 10.1056/NEJM200002103420603

2. Nathan DM, Cleary PA, Backlund JY, Genuth SM, Lachin JM, Orchard TJ, et al.; Diabetes Control and Complications Trial/Epidemiology of Diabetes Interventions and Complications (DCCT/EDIC) Study Research Group. Intensive diabetes treatment and cardiovascular disease in patients with type 1 diabetes. N Engl J Med. 2005 Dec;353(25):2643–53. 10.1056/NEJMoa052187

3. Chehregosha H, Khamseh ME, Malek M, Hosseinpanah F, Ismail-Beigi F. A View Beyond HbA1c: Role of Continuous Glucose Monitoring. Diabetes Ther. 2019 Jun;10(3):853–63. 10.1007/s13300-019-0619-1

4. Beck RW, Riddlesworth TD, Ruedy K, Ahmann A, Haller S, Kruger D, et al.; DIAMOND Study Group. Continuous Glucose Monitoring Versus Usual Care in Patients With Type 2 Diabetes Receiving Multiple Daily Insulin Injections: A Randomized Trial. Ann Intern Med. 2017 Sep;167(6):365–74. 10.7326/M16-2855

5. Lind M, Polonsky W, Hirsch IB, Heise T, Bolinder J, Dahlqvist S, et al. Continuous Glucose Monitoring vs Conventional Therapy for Glycemic Control in Adults With Type 1 Diabetes Treated With Multiple Daily Insulin Injections: The GOLD Randomized Clinical Trial. JAMA. 2017 Jan;317(4):379–87. 10.1001/jama.2016.19976