The purpose of this study was to develop a three-dimensional (3D) motion analysis based anatomical wrist joint coordinate system for measurement of in-vivo wrist kinematics. The convergent validity and reliability of the 3D motion analysis implementation was quantified and compared with manual and electrogoniometry techniques on 10 cadaveric specimens. Fluoroscopic measurements were used as the reference. The 3D motion analysis measurements (mean absolute difference [MAD] = 3.6°) were significantly less different (P < .005) than manual goniometry (MAD = 5.7°) but not (P = .066, power = 0.45) electrogoniometry (MAD = 5.0°) compared with fluoroscopy. The intraclass correlation coefficient (ICC[2,1]) was highest for 3D motion analysis compared with manual and electrogoniometry, suggesting better reliability for this technique. To demonstrate the utility of this new wrist joint coordinate system, normative data from 10 healthy subjects was obtained while throwing a dart.