Abstract

AbstractThis paper describes cooperation of the Slovak University of Technology in Bratislava (STU) and the Slovak Institute of Metrology (SMU) in the verification of the calibration factors for the solid-state nuclear track detector utilized for radon dosimetry and establishing traceability for the STU laboratory. The SMU operates the Air Radon Standard which consists of a radon chamber and calibrated radon atmosphere monitoring system. A specification of the radon chamber is that the radon concentration changes during the exposure time. The radon atmosphere is created by the insertion of a specified amount of radon gas into the chamber, which subsequently decays over time. The STU laboratory is equipped with the TASLImage™ system for radon dosimetry with corresponding TASTRAK CR-39 type detectors and diffusion chambers. These chambers allow radon to diffuse inside and prevent dust and radon progeny to enter the container. Each diffusion container is characterized by its own diffusion rate, which should be considered in the metrology calibration approach, especially when the radon chamber is operated in the decay mode. Therefore, several measurements were carried out to determine the optimal condition for conversion factor verification. The experiments were conducted for low radon concentrations, but still in the range that the radon chamber can reliably provide (100 Bq m−3 to 100 kBq m−3) and in combination with several times of exposure. Based on the achieved results the default calibration factor was verified, and an optimal calibration approach was proposed. To improve the capability of the radon measurement, a custom diffusion container design was proposed. The design incorporates compact dimensions with an aerosol filter at the entrance to the container. This allows to increase in an entry hole and thus increases the diffusion rate. Subsequently, the homogeneity of the radon atmosphere in the chamber was tested.

Funder

Agentúra na Podporu Výskumu a Vývoja

Agentúra Ministerstva Školstva, Vedy, Výskumu a Športu SR

Slovak Technical University

Publisher

Springer Science and Business Media LLC

Subject

Physical and Theoretical Chemistry,General Physics and Astronomy,General Materials Science

Reference8 articles.

1. F. Bochicchio et al., Int. J. Environ. Res. Public Health 19, 4114 (2022)

2. National action plan § 135, Law num. 87/2018 on radiation protection

3. P. Blahušiak et al., Rad. Prot. Dos. 191, 160 (2020)

4. TASLIMAGE Radon and Neutron Dosimetry System—A Complete System for Etching, Scanning and Analysing TASTRAK Plastic Nuclear Track Detector. Track Analysis Systems Ltd, Bristol (2020).

5. B. Tate, S. Iong, Acceptance Testing of the TASL Radon Dosimetry System (Australian Radiation Protection and Nuclear Safety Agency, Australia, 2016)

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3