CR-39 detector-based Radon dosimetry system calibration in the self-decay mode
-
Published:2023-06-02
Issue:10
Volume:232
Page:1493-1500
-
ISSN:1951-6355
-
Container-title:The European Physical Journal Special Topics
-
language:en
-
Short-container-title:Eur. Phys. J. Spec. Top.
Author:
Lüley J.ORCID,
Filova V.,
Blahušiak P.,
Vrban B.,
Čerba Š.,
Bonková I.,
Nečas V.
Abstract
AbstractThis paper describes cooperation of the Slovak University of Technology in Bratislava (STU) and the Slovak Institute of Metrology (SMU) in the verification of the calibration factors for the solid-state nuclear track detector utilized for radon dosimetry and establishing traceability for the STU laboratory. The SMU operates the Air Radon Standard which consists of a radon chamber and calibrated radon atmosphere monitoring system. A specification of the radon chamber is that the radon concentration changes during the exposure time. The radon atmosphere is created by the insertion of a specified amount of radon gas into the chamber, which subsequently decays over time. The STU laboratory is equipped with the TASLImage™ system for radon dosimetry with corresponding TASTRAK CR-39 type detectors and diffusion chambers. These chambers allow radon to diffuse inside and prevent dust and radon progeny to enter the container. Each diffusion container is characterized by its own diffusion rate, which should be considered in the metrology calibration approach, especially when the radon chamber is operated in the decay mode. Therefore, several measurements were carried out to determine the optimal condition for conversion factor verification. The experiments were conducted for low radon concentrations, but still in the range that the radon chamber can reliably provide (100 Bq m−3 to 100 kBq m−3) and in combination with several times of exposure. Based on the achieved results the default calibration factor was verified, and an optimal calibration approach was proposed. To improve the capability of the radon measurement, a custom diffusion container design was proposed. The design incorporates compact dimensions with an aerosol filter at the entrance to the container. This allows to increase in an entry hole and thus increases the diffusion rate. Subsequently, the homogeneity of the radon atmosphere in the chamber was tested.
Funder
Agentúra na Podporu Výskumu a Vývoja
Agentúra Ministerstva Školstva, Vedy, Výskumu a Športu SR
Slovak Technical University
Publisher
Springer Science and Business Media LLC
Subject
Physical and Theoretical Chemistry,General Physics and Astronomy,General Materials Science
Reference8 articles.
1. F. Bochicchio et al., Int. J. Environ. Res. Public Health 19, 4114 (2022)
2. National action plan § 135, Law num. 87/2018 on radiation protection
3. P. Blahušiak et al., Rad. Prot. Dos. 191, 160 (2020)
4. TASLIMAGE Radon and Neutron Dosimetry System—A Complete System for Etching, Scanning and Analysing TASTRAK Plastic Nuclear Track Detector. Track Analysis Systems Ltd, Bristol (2020).
5. B. Tate, S. Iong, Acceptance Testing of the TASL Radon Dosimetry System (Australian Radiation Protection and Nuclear Safety Agency, Australia, 2016)
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献