Abstract
AbstractThe evaluation of minimum detectable activity (MDA) for a radionuclide in a gamma-ray spectrum is generally carried out through the computation of a suitable background count. This task is sometimes difficult for complex spectra for the presence of many photopeaks which make the trend of continuum extremely variable due to multiple dispersion effects and interference factors. It follows that the MDA assessment must be take into account the contributions of all gamma emissions of radionuclides contained in a sample and its value can be significantly higher than that determined by considering only the background of the spectrometric system due to the overlapping of other peaks. A procedure or an algorithm to determine, each time, the count values to be used for the calculation of MDA is interesting and useful. In this work, some of the more recent algorithms proposed for background subtraction in a gamma-ray spectrum have been examined, applying them in an inverse way for the evaluation of baseline trend in the whole energy range. Among the algorithms examined, particular attention was paid to the application of SNIP (statistical sensitive nonlinear iterative peak clipping) algorithms, which are the simplest to adopt and implement in an application procedure. The results obtained in the analysis of test gamma-ray spectra are satisfactory and allow to quickly determine the MDA values with a formulation based on the ISO-11929 standard.
Funder
Università degli Studi di Palermo
Publisher
Springer Science and Business Media LLC
Subject
General Physics and Astronomy,Fluid Flow and Transfer Processes
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献