Suture-Button versus Screw Fixation in a Syndesmosis Rupture Model: A Biomechanical Comparison

Author:

Soin Sandeep P.1,Knight Trevor A.1,Dinah A. Feroz1,Mears Simon C.1,Swierstra Bart A.1,Belkoff Stephen M.1

Affiliation:

1. Baltimore, MD

Abstract

Background: Flexible suture-button devices are used to stabilize the distal tibiofibular joint after syndesmotic rupture, but little is known about how they affect fibular motion during cyclic physiological loading. This study examined how much fibular motion occurs during cyclic loading of ankles repaired with suture-buttons or screws relative to one with intact ligaments. Materials and Methods: Ten pairs of cadaveric legs with intact ligaments were tested in axial compression (750 N), external rotational torque (7.5 Nm), and a combination thereof. Fibular rotation and translation in the sagittal and frontal planes were measured with linear variable displacement transducers. Then we sectioned the anterior tibiofibular, posterior tibiofibular, deltoid, and interosseous ligaments. Ankles were repaired with suture-buttons or a 3.5-mm screw through four-cortices. The ankles were tested for 10,000 cycles using the same loading protocol. Surviving ankles were brought to failure by increasing external rotation (1 degree/sec). Data from the linear variable differential transducers were used to calculate fibular motion at 100, 1,000, 5,000, and 10,000 cycles. Results: There was no significant difference in the effect of the suture-button or screw for translation in the sagittal or coronal plane or for rotation about the fibular axis. The screw repair had a significantly greater ( p = 0.02) failure torque (median, 26.5 Nm; inner quartile range, 25.7 to 35.2 Nm) than did the suture-button repair (median, 23.6 Nm; inner quartile range, 16.5 to 25.6 Nm). Conclusion: The fibular motion that occurs during cyclic loading appears to be similar for the suture-button and syndesmotic screw. Motion in both constructs was similar, but neither restored native ankle motion. Clinical Relevance: It appears the suture-button behaves similarly to the syndesmotic screw in the syndesmotic rupture injury model tested. Clinical trials are needed to determine how the device performs in vivo.

Publisher

SAGE Publications

Subject

Orthopedics and Sports Medicine,Surgery

Cited by 77 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3