The Importance of Temporal Resolution for Ultra-High-Resolution Coronary Angiography

Author:

Mergen Victor,Sartoretti Thomas,Cundari Giulia,Serifovic Medina,Higashigaito Kai,Allmendinger Thomas,Schmidt Bernhard,Flohr Thomas,Manka Robert,Eberhard Matthias,Alkadhi Hatem

Abstract

Purpose The aim of this study was to assess the effect of temporal resolution on subjective and objective image quality of coronary computed tomography angiography (CCTA) in the ultra-high-resolution (UHR) mode with dual-source photon-counting detector (PCD) CT. Materials and Methods This retrospective, institutional review board–approved study evaluated 30 patients (9 women; mean age, 80 ± 10 years) undergoing UHR CCTA with a clinical dual-source PCD-CT scanner. Images were acquired with a tube voltage of 120 kV and using a collimation of 120 × 0.2 mm. Gantry rotation time was 0.25 seconds. Each scan was reconstructed using both single-source and dual-source data resulting in an image temporal resolution of 125 milliseconds and 66 milliseconds, respectively. The average heart rate and the heart rate variability were recorded. Images were reconstructed with a slice thickness of 0.2 mm, quantum iterative reconstruction strength level 4, and using the Bv64 and Bv72 kernel for patients without and with coronary stents, respectively. For subjective image quality, 2 experienced readers rated motion artifacts and vessel delineation, or in-stent lumen visualization using 5-point discrete visual scales. For objective image quality, signal-to-noise ratio, contrast-to-noise ratio, stent blooming artifacts, and vessel and stent sharpness were quantified. Results Fifteen patients had coronary stents, and 15 patients had no coronary stents. The mean heart rate and heart rate variability during data acquisition were 72 ± 10 beats per minute and 5 ± 6 beats per minute, respectively. Subjective image quality in the right coronary artery, left anterior descending, and circumflex artery was significantly superior in 66 milliseconds reconstructions compared with 125 milliseconds reconstructions for both readers (all P's < 0.01; interreader agreement, Krippendorff α = 0.84–1.00). Subjective image quality deteriorated significantly at higher heart rates for 125 milliseconds (ρ = 0.21, P < 0.05) but not for 66 milliseconds reconstructions (ρ = 0.11, P = 0.22). No association was found between heart rate variability and image quality for both 125 milliseconds (ρ = 0.09, P = 0.33) and 66 milliseconds reconstructions (ρ = 0.13, P = 0.17), respectively. Signal-to-noise ratio and contrast-to-noise ratio were similar between 66 milliseconds and 125 milliseconds reconstructions (both P's > 0.05), respectively. Stent blooming artifacts were significantly lower on 66 milliseconds than on 125 milliseconds reconstructions (46.7% ± 10% vs 52.9% ± 8.9%, P < 0.001). Higher sharpness was found in 66 milliseconds than in 125 milliseconds reconstructions both in native coronary arteries (left anterior descending artery: 1031 ± 265 ∆HU/mm vs 819 ± 253 ∆HU/mm, P < 0.01; right coronary artery: 884 ± 352 ∆HU/mm vs 654 ± 377 ∆HU/mm, P < 0.001) and stents (5318 ± 3874 ∆HU/mm vs 4267 ± 3521 ∆HU/mm, P < 0.001). Conclusions Coronary angiography with PCD-CT in the UHR mode profits considerably from a high temporal resolution, resulting in less motion artifacts, superior vessel delineation and in-stent lumen visualization, less stent blooming artifacts, and superior vessel and stent sharpness.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Radiology, Nuclear Medicine and imaging,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3