Photon-Counting Detector CT Radiological-Histological Correlation in Cadaveric Human Lung Nodules and Airways

Author:

Hata AkinoriORCID,Yanagawa MasahiroORCID,Ninomiya Keisuke,Kikuchi Noriko,Kurashige Masako,Masuda Chiaki,Yoshida Tsubasa,Nishigaki Daiki,Doi Shuhei,Yamagata Kazuki,Yoshida Yuriko,Ogawa Ryo,Tokuda Yukiko,Morii EiichiORCID,Tomiyama NoriyukiORCID

Abstract

Objectives The aim of this study was to compare the performances of photon-counting detector computed tomography (PCD-CT) and energy-integrating detector computed tomography (EID-CT) for visualizing nodules and airways in human cadaveric lungs. Materials and Methods Previously obtained 20 cadaveric lungs were scanned, and images were prospectively acquired by EID-CT and PCD-CT at a radiation dose with a noise level equivalent to the diagnostic reference level. PCD-CT was scanned with ultra-high-resolution mode. The EID-CT images were reconstructed with a 512 matrix, 0.6-mm thickness, and a 350-mm field of view (FOV). The PCD-CT images were reconstructed at 3 settings: PCD-512: same as EID-CT; PCD-1024-FOV350: 1024 matrix, 0.2-mm thickness, 350-mm FOV; and PCD-1024-FOV50: 1024 matrix, 0.2-mm thickness, 50-mm FOV. Two specimens per lung were examined after hematoxylin and eosin staining. The CT images were evaluated for nodules on a 5-point scale and for airways on a 4-point scale to compare the histology. The Wilcoxon signed rank test with Bonferroni correction was performed for statistical analyses. Results Sixty-seven nodules (1321 μm; interquartile range [IQR], 758–3105 μm) and 92 airways (851 μm; IQR, 514–1337 μm) were evaluated. For nodules and airways, scores decreased in order of PCD-1024-FOV50, PCD-1024-FOV350, PCD-512, and EID-CT. Significant differences were observed between series other than PCD-1024-FOV350 versus PCD-1024-FOV50 for nodules (PCD-1024-FOV350 vs PCD-1024-FOV50, P = 0.063; others P < 0.001) and between series other than EID-CT versus PCD-512 for airways (EID-CT vs PCD-512, P = 0.549; others P < 0.005). On PCD-1024-FOV50, the median size of barely detectable nodules was 604 μm (IQR, 469–756 μm) and that of barely detectable airways was 601 μm (IQR, 489–929 μm). On EID-CT, that of barely detectable nodules was 837 μm (IQR, 678–914 μm) and that of barely detectable airways was 1210 μm (IQR, 674–1435 μm). Conclusions PCD-CT visualized small nodules and airways better than EID-CT and improved with high spatial resolution and potentially can detect submillimeter nodules and airways.

Publisher

Ovid Technologies (Wolters Kluwer Health)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3