Diagnostic Performance and Therapeutic Consequence of Thromboelastometry Activated by Kaolin versus  a Panel of Specific Reagents

Author:

Larsen Ole Halfdan1,Fenger-Eriksen Christian2,Christiansen Kirsten3,Ingerslev Jørgen4,Sørensen Benny5

Affiliation:

1. Ph.D. Student, Centre for Haemophilia and Thrombosis, Aarhus University Hospital, Skejby, Denmark.

2. Fellow Anesthesiologist, Department of Anesthesiology, Aarhus University Hospital.

3. Research Technologist, Centre for Haemophilia and Thrombosis, Aarhus University Hospital.

4. Professor, Centre for Haemophilia and Thrombosis, Aarhus University Hospital.

5. Associate Professor, Centre for Haemophilia and Thrombosis, Aarhus University Hospital, Director and Honorary Lecturer, Haemostasis Research Unit, Centre for Haemostasis and Thrombosis, St. Thomas' Hospital, London, United Kingdom.

Abstract

Background Thromboelastography/metry (TEG®; Haemoscope, Niles, IL/ROTEM®; Tem International GmbH, Munich, Germany) is increasingly used to guide transfusion therapy. This study investigated the diagnostic performance and therapeutic consequence of using kaolin-activated whole blood compared with a panel of specific TEM®-reagents to distinguish: dilutional coagulopathy, thrombocytopenia, hyperfibrinolysis, and heparinization. Methods Blood was drawn from 11 healthy volunteers. Dilutional coagulopathy was generated by 50% dilution with hydroxyethyl starch 130/0.4 whereas thrombocytopenia (mean platelet count 20 ×10⁹/l) was induced using a validated model. Hyperfibrinolysis and heparin contamination were generated by tissue plasminogen activator 2 nM and unfractionated heparin 0.1U/ml, respectively. Coagulation tests were run on ROTEM® delta. Results Kaolin-activated whole blood showed no differences between dilutional coagulopathy and thrombocytopenia (mean clotting time 450 s vs. 516 s, α-angle 47.1° vs. 41.5°, maximum clot firmness 35.0 mm vs. 34.2 mm, all P values ≥0.14). Hyperfibrinolysis specifically disclosed an increased maximum lysis (median: 100%, all P values less than 0.001), and heparin induced a distinctly prolonged clotting time (2283 s, all P values less than 0.02). The coagulopathies were readily distinguishable using a panel of TEM-reagents. In particular, dilutional coagulopathy was separated from thrombocytopenia using FIBTEM (maximum clot firmness 1.9 mm vs. 11.2 mm, P < 0.001). The run time of analysis to achieve diagnostic data was shorter applying a panel of TEM-reagents. A transfusion algorithm based on kaolin suggested platelets in case of dilutional coagulopathy, whereas an algorithm applying TEM-reagents suggested fibrinogen. Conclusion Monoanalysis with kaolin was unable to distinguish coagulopathies caused by dilution from that of thrombocytopenia. Algorithms based on the use of kaolin may lead to unnecessary transfusion with platelets, whereas the application of TEM-reagents may result in goal-directed fibrinogen substitution.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Anesthesiology and Pain Medicine

Reference44 articles.