Abstract
Introduction:
This report describes a case of positive dysphotopsia (PD) after refractive lens exchange (RLE) with implantation of an extended depth-of-focus (EDOF) intraocular lens (IOL).
Patient and Clinical Findings:
A healthy 48-year-old man underwent RLE with bilateral implantation of a nontoric EDOF IOL and suffered from monocular triplopia in both eyes. The visual acuity was suboptimal with an uncorrected distance visual acuity of 20/25 in the right eye and 20/30 in the left eye.
Diagnosis, Intervention and Outcomes:
Slitlamp examination showed well-centered implanted IOLs, without capsular distension syndrome. Bilateral IOL exchange was performed with an aspheric monofocal IOL with a 7.0 mm optic diameter, which diminished the PD complaints.
Conclusions:
This case report describes disturbing PD after bilateral implantation of an EDOF IOL. Exchange of these EDOF IOLs with a larger optical diameter monofocal IOL resulted in resolution of symptoms.
Publisher
Ovid Technologies (Wolters Kluwer Health)