Comparison of 1-year Follow-up Echocardiographic Outcomes of Sapien 3 Versus Evolut R Bioprosthetic Transcatheter Aortic Valves: A Single-center Retrospective Iranian Cohort Study

Author:

Sahebjam Mohammad12,Zoroufian Arezou12,Hajizeinali Alimohammad1,Salarifar Mojtaba1,Jalali Arash1,Ayati Aryan1,Farmanesh Mahkameh12

Affiliation:

1. Cardiovascular Diseases Research Institute, Tehran University of Medical Sciences, Tehran, Iran

2. Department of Echocardiography, Tehran Heart Center, Tehran University of Medical Sciences, Tehran, Iran.

Abstract

Objective: The current study aimed to compare 1-year echocardiographic outcomes of the new generations of self-expanding (Evolut R) versus balloon-expandable (Sapien 3) bioprosthetic transcatheter aortic valves. Methods: In this study, gradients and flow velocities obtained from transthoracic Doppler-echocardiography were retrospectively collected from patients who underwent 2 new generations of transcatheter aortic valve implantation interventions with Sapien 3 and Evolut R valves. Patients underwent echocardiography before the procedure and at discharge, 6 months, and 1-year follow-up. Results: Of the 66 patients, 28 received Sapien 3 and 38 received Evolut R valves. Evolut R valve presented a lower mean gradient at all follow-up time points compared with Sapien 3 valves (14.4 mm Hg, 14.9 mm Hg, 15.5 mm Hg compared with 10.1 mm Hg, 11.6 mm Hg, 11.8 mm Hg, respectively; all P-values <0.001). Small valve sizes of Evolut R, including 23 and 26, had higher echocardiographic mean gradient or peak gradient at the time of discharge compared with larger valves, including sizes 29 and 34 (11.1 mm Hg and 11.2 mm Hg vs. 10.2 mm Hg, 9.1 mm Hg) and 1-year follow-up (11.0 mm Hg, 11.0 mm Hg vs. 9.9 mm Hg, 8.4 mm Hg; all P-values = 0.001). Although Sapien 3 valves demonstrated a higher peak gradient in smaller sizes at discharge (18.44 mm Hg in size 23 vs. 17.9 mm Hg, 16.5 mm Hg in size 26 and 29, respectively; P = 0.001), the peak gradients did not show a statistically significant difference in the 1-year follow-up. Conclusions: The current study detected significantly lower mean and peak gradients in Evolut R compared with Sapien 3 at all follow-up time points. Furthermore, smaller valve sizes were associated with significantly higher gradients at all follow-ups, regardless of the valve type.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3