The Feasibility of the Three-Dimensional Footswitch-Operated Robotic Arm Exoscope for Cochlear Implant Surgery

Author:

Karinen Eerika1,Iso-Mustajärvi Matti2,Dietz Aarno

Affiliation:

1. Department of Otorhinolaryngology, Satasairaala Hospital, Pori, Finland

2. Department of Otorhinolaryngology, Kuopio University Hospital, Kuopio, Finland

Abstract

Objective To compare the three-dimensional (3D) footswitch-operated robotic arm exoscope with the operating microscope (OM) in cochlear implant surgery. Study design Matched case–control study. Patients Cochlear implantation was performed with the exoscope on unselected patients with normal temporal anatomy. The control group that underwent cochlear implantation with the OM was case matched with respect to age, anatomy, surgical technique, and type of anesthesia. Interventions Cochlear implantation performed with the 3D exoscope. Main Outcome measures Surgical time, occupation of the operation theater, surgical results, and user experience evaluated by a questionnaire. Results Eleven patients (13 ears) were successfully operated on with the exoscope. In the exoscope group, we observed one minor intraoperative complication, where the middle dura was exposed during mastoidectomy. Although no clear preference was evident for either device in the overall rating, the subdomain rating revealed that the exoscope’s image quality was deemed inferior, especially at higher magnifications where pixelation became noticeable. The exoscope received higher scores for usability, particularly excelling in terms of surgeon’s ergonomic and comfortability. There was a statistically significant difference in mean surgical time, 146 and 129 min for the exoscope and OM group, respectively. Conclusions Cochlear implant surgery was found to be feasible with a 3D exoscope. However, there is a learning curve to overcome regarding handling and the different quality of the image. The exoscope provides better ergonomics for the surgeon.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Neurology (clinical),Sensory Systems,Otorhinolaryngology

Reference16 articles.

1. The history of the microscope for use in ear surgery;Am J Otol,2000

2. The risks of being otologist, an ergonomic and occupational hazard review;Otol Neurotol,2020

3. Surgical ergonomics;Surgeons as Educators Published online,2018

4. Advances in intraoperative optics: a brief review of current exoscope platforms;Oper Neurosurg (Hagerstown),2020

5. The exoscope in neurosurgery: an innovative “point of view”. A systematic review of the technical, surgical, and educational aspects;World Neurosurg,2019

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Exoscope-assisted surgery in otology and neurotology;Current Opinion in Otolaryngology & Head & Neck Surgery;2024-07-24

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3