Abstract
SIGNIFICANCE
New refractive technologies are consistently emerging in the optometry market, necessitating validation against current clinical standards.
PURPOSE
This study aimed to compare the refractive measurements between standard digital phoropter refraction and the Chronos binocular refraction system.
METHODS
Standardized subjective refraction was conducted on 70 adult participants using two separate refraction systems. The final subjective values from both devices were compared for M, J
0, and J
45. The time taken to complete refraction and patient's comfort were also evaluated.
RESULTS
Good agreement was found between the standard and Chronos refraction, with narrow mean differences (including 95% confidence intervals) and no significant bias for M (0.03 D, −0.05 to 0.11 D), J
0 (−0.02 D, −0.05 to −0.01 D), and J
45 (−0.01 D, −0.03 to 0.01 D). The bounds of the limits of agreement of M were −0.62 (lower bound; −0.76 to −0.49) and 0.68 (upper bound; 0.54 to 0.81), those of J
0 were −0.24 (lower bound; −0.29 to −0.19) and 0.19 (upper bound; 0.15 to 0.24), and those of J
45 were −0.18 (lower bound; −0.21 to −0.14) and 0.16 (upper bound; 0.12 to 0.19). No significant differences were noted between the two techniques for any of the refraction components (M standard = −3.03 ± 2.42 D, M novel = −3.06 ± 2.37 D, z = 0.07, P = .47; J
0 standard = 0.12 ± 0.40 D, J
0 novel = 0.15 ± 0.41 D, z = 1.32, P = .09; J
45 standard = −0.04 ± 0.19 D, J
45 novel = −0.03 ± 0.19 D, z = 0.50, P = .31). The Chronos was significantly faster than the standard technique, with an average difference of 19 seconds (standard, 190 ± 44 seconds; novel, 171 ± 38 seconds; z = 4.91; P < .001).
CONCLUSIONS
The final subjective refraction end points of the standard technique and the Chronos were well aligned in this group of adult participants, and no statistically or clinically significant differences were noted in M, J
0, or J
45 components. The Chronos offered improved efficiency, meeting the demands of eye care.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Reference28 articles.
1. Clinical Evaluation of the Allergan Humphrey 500 Autorefractor and the Nidek AR-1000 Autorefractor;Br J Ophthalmol,1996
2. The Repeatability of Automated and Clinician Refraction;Optom Vis Sci,1998
3. Repeatability of Subjective and Objective Refraction;Optom Vis Sci,1995
4. Steps Towards Smarter Solutions in Optometry and Ophthalmology—Inter-device Agreement of Subjective Methods to Assess the Refractive Errors of the Eye;Healthcare (Basel),2016
5. Beyond Traditional Subjective Refraction;Curr Opin Ophthalmol,2022