Affiliation:
1. From the Department of Anesthesiology (A.J., A.D.), and the Department of Intensive Care (F.S., J.C., J.-L.V.), Erasme University Hospital, Université Libre de Bruxelles, Brussels, Belgium; the Department of Anesthesiology and Intensive Care, Hôpitaux Universitaires Paris-Sud, Université Paris-Sud, Université Paris-Saclay, Hôpital De Bicêtre, Assistance Publique Hôpitaux de Paris, Le Kremlin-Bicê
Abstract
Abstract
Editor’s Perspective
What We Already Know about This Topic
What This Article Tells Us That Is New
Background
Multiple studies have reported associations between intraoperative hypotension and adverse postoperative complications. One of the most common interventions in the management of hypotension is vasopressor administration. This approach requires careful and frequent vasopressor boluses and/or multiple adjustments of an infusion. The authors recently developed a closed-loop controller that titrates vasopressors to maintain mean arterial pressure (MAP) within set limits. Here, the authors assessed the feasibility and overall performance of this system in a swine model. The authors hypothesized that the closed-loop controller would be able to maintain MAP at a steady, predefined target level of 80 mmHg for greater than 85% of the time.
Methods
The authors randomized 14 healthy anesthetized pigs either to a control group or a closed-loop group. Using infusions of sodium nitroprusside at doses between 65 and 130 µg/min, we induced four normovolemic hypotensive challenges of 30 min each. In the control group, nothing was done to correct hypotension. In the closed-loop group, the system automatically titrated norepinephrine doses to achieve a predetermined MAP of 80 mmHg. The primary objective was study time spent within ±5 mmHg of the MAP target. Secondary objectives were performance error, median performance error, median absolute performance error, wobble, and divergence.
Results
The controller maintained MAP within ±5 mmHg of the target for 98 ± 1% (mean ± SD) of the time. In the control group, the MAP was 80 ± 5 mmHg for 14.0 ± 2.8% of the time (P< 0.0001). The MAP in the closed-loop group was above the target range for 1.2 ± 1.2% and below it for 0.5 ± 0.9% of the time. Performance error, median performance error, median absolute performance error, wobble, and divergence were all optimal.
Conclusions
In this experimental model of induced normovolemic hypotensive episodes in pigs, the automated controller titrated norepinephrine infusion to correct hypotension and keep MAP within ±5 mmHg of target for 98% of management time.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Anesthesiology and Pain Medicine
Cited by
41 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献