Comparison of a 2.7-mm and 3.5-mm locking compression plate for ulnar fractures: a biomechanical evaluation

Author:

Wahbeh Jenna M.12ORCID,Kelley Benjamin V.3ORCID,Shokoohi Cyrus2,Park Sang-Hyun13,Devana Sai K.3,Ebramzadeh Edward13ORCID,Sangiorgio Sophia N.123,Jeffcoat Devon M.3ORCID

Affiliation:

1. The J. Vernon Luck, Sr., M.D. Orthopaedic Research Center, Luskin Orthopaedic Institute for Children in Alliance with UCLA, Los Angeles, CA

2. University of California, Los Angeles, Department of Bioengineering, Los Angeles, CA

3. University of California, Los Angeles, Department of Orthopaedic Surgery, Los Angeles, CA

Abstract

Abstract Objectives: Implant prominence after ulnar fracture fixation may be mitigated by the use of lower profile plates. The biomechanical strength and stability of 2.7-mm and 3.5-mm locking compression plates for fixation were compared. Methods: Two fracture conditions, transverse (N = 10) and oblique (N = 10), were evaluated in an in vitro study. Half of the specimens for each condition were fixed with 2.7-mm plates and the other half with 3.5-mm plates, all fixed with conventional dynamic compression mechanisms. Specimens were loaded under ±2 Nm of cyclic axial torsion, then under 10 Nm of cyclic cantilever bending, and bending to failure. Interfragmentary motion and strain were analyzed to determine construct stability as a function of fracture pattern and plate size. Results: Interfragmentary motion was significantly larger in all constructs fixed with 2.7-mm plates, compared with 3.5-mm plates (P < 0.01). The 2.7-mm constructs with transverse fractures had the greatest motion, ranging between 5° and 10° under axial rotation and 5.0–6.0 mm under bending. Motions were the lowest for 3.5-mm constructs with oblique fractures, ranging between 3.2 and 4.2 mm under bending and 2°–3.5° for axial rotation. For oblique fractures, the bending moment at ultimate failure was 31.4 ± 3.6 Nm for the 2.7-mm constructs and 10.0 ± 1.9 Nm for 3.5-mm constructs (P < 0.01). Similarly, for transverse fractures, the bending moment was 17.9 ± 4.0 Nm for the 2.7-mm constructs and 9.7 ± 1.3 Nm for the 3.5-mm constructs (P < 0.01). Conclusions: Although 3.5-mm plates were more effective at reducing fracture motion, they were consistently associated with refracture at the distal-most screw hole under load to failure. By contrast, 2.7-mm plates plastically deformed despite excessive loads, potentially avoiding a subsequent fracture. Level of Evidence: Level V.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

General Medicine

Reference54 articles.

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3