Single-use Lens Compromise During Selective Laser Trabeculoplasty

Author:

Liu Alice S.1,Lyons Lance J.2,Khanna Cheryl L.2,Roddy Gavin W.2,Sit Arthur J.2

Affiliation:

1. Mayo Clinic Alix School of Medicine, Rochester, MN, USA

2. Department of Ophthalmology, Mayo Clinic, Rochester, MN, USA

Abstract

Precis: Damage to disposable selective laser trabeculoplasty (SLT) lenses correlated with the amount of total laser energy. Furthermore, greater lens damage was associated with diminished patient treatment response. Purpose: Disposable selective laser trabeculoplasty (SLT) lenses have been observed to be damaged during use by the laser, potentially affecting therapeutic response. This study sought to identify factors associated with the magnitude of lens damage and its effect on treatment outcomes. Patient and Methods: We analyzed 113 eyes from 82 patients who underwent SLT between 2020-2021 at an American academic medical center. For each procedure, we recorded baseline patient characteristics, procedure settings, physician age (as a possible factor causing laser defocus due to accommodation), and area of lens damage. Treatment response was calculated as the difference between pre- and post-operative intraocular pressure (IOP). Results: Area of lens damage was associated with greater total laser energy (r=0.34, P<0.001) and greater mean energy per application (r=0.37, P<0.001). IOP reduction decreased with increasing lens damage (r=−0.19, P=0.022). Lens damage was not associated with patient characteristics indicative of glaucoma severity or user settings (physician age, slit lamp used). Greater IOP reduction was associated with higher pre-op IOP (r=0.46, P<0.001) and a smaller cup-to-disc ratio (r=0.22, P=0.036). IOP reduction was not associated with glaucoma severity, patient age, or the number of glaucoma medications. Conclusions: Increased laser damage to disposable SLT lenses was associated with diminished treatment effect. The amount of damage was correlated with total and average laser energy, but not patient or physician factors. Overall, these findings call for reconsideration of using disposable SLT lenses as part of routine practice.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Ophthalmology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3