Affiliation:
1. Department of Anaesthesiology
2. Division of Paediatric Surgery, Department of General, Visceral, and Transplantation Surgery
3. Paediatric Neurology and Metabolic Medicine, Centre for Paediatrics and Adolescent Medicine, Heidelberg University, Medical Faculty Heidelberg, Heidelberg, Germany
Abstract
Background
In hypoxemic children with difficult airway, or for minor elective procedures, the use of a supraglottic airway device may be preferred to endotracheal intubation, whether with a laryngeal mask or laryngeal tube. Second-generation laryngeal masks may offer a better safety profile. Whether they should be preferred to laryngeal tubes is unknown. This study aimed to compare the efficacy and safety of second-generation laryngeal masks and laryngeal tubes in children.
Methods
This randomised-controlled trial was conducted in a single university hospital in children <18 years undergoing elective anaesthesia in urology, minor paediatric surgery and gynaecology. Patients were 1 : 1 randomised to the laryngeal mask or laryngeal tube group. Children were allocated a second-generation laryngeal tube or a second-generation laryngeal mask as the primary airway device. The primary endpoint was insertion time. Secondary endpoints included first-attempt success, overall success and complications, which included hypoxia (SpO2 < 90%), laryngospasm, bronchospasm, aspiration and bleeding.
Results
In total, 135 patients were randomised, with 61 allocated to the laryngeal tube and 74 to the laryngeal mask group, with a median age of 5.4 and 4.9 years, respectively. Median insertion time was significantly longer in the laryngeal tube group (37 vs. 31 s; difference of medians: 6.0 s; 95% confidence interval: 0.0–13.0). The laryngeal tube had a significantly lower first-attempt (41.0%) and overall success rate (45.9%) than the laryngeal mask (90.5% and 97.3%, respectively). Those allocated to the laryngeal tube group had a higher ratio of complications (27.8%) compared to the laryngeal mask group (2.7%).
Conclusion
This randomised-controlled trial reported that in children undergoing elective anaesthesia, the use of a laryngeal tube was associated with a longer insertion time.
Publisher
Ovid Technologies (Wolters Kluwer Health)