Combined Platelet and Erythrocyte Salvage: Evaluation of a New Filtration-based Autotransfusion Device

Author:

Mansour Alexandre,Decouture Benoit,Roussel Mikaël,Lefevre Charles,Skreko Lucie,Picard Véronique,Ouattara Alexandre,Bachelot-Loza Christilla,Gaussem Pascale,Nesseler Nicolas,Gouin-Thibault Isabelle

Abstract

Background The SAME device (i-SEP, France) is an innovative filtration-based autotransfusion device able to salvage and wash both red blood cells and platelets. This study evaluated the device performances using human whole blood with the hypothesis that the device will be able to salvage platelets while achieving a erythrocyte yield of 80% and removal ratios of 90% for heparin and 80% for major plasma proteins without inducing signification activation of salvaged cells. Methods Thirty healthy human whole blood units (median volume, 478 ml) were diluted, heparinized, and processed by the device in two consecutive treatment cycles. Samples from the collection reservoir and the concentrated blood were analyzed. Complete blood count was performed to measure blood cell recovery rates. Flow cytometry evaluated the activation state and function of platelets and leukocytes. Heparin and plasma proteins were measured to assess washing performance. Results The global erythrocyte yield was 88.1% (84.1 to 91.1%; median [25th to 75th]) with posttreatment hematocrits of 48.9% (44.8 to 51.4%) and 51.4% (48.4 to 53.2%) for the first and second cycles, respectively. Ektacytometry did not show evidence of erythrocyte alteration. Platelet recovery was 36.8% (26.3 to 43.4%), with posttreatment counts of 88 × 109/l (73 to 101 × 109/l) and 115 × 109/l (95 to 135 × 109/l) for the first and second cycles, respectively. Recovered platelets showed a low basal P-selectin expression at 10.8% (8.1 to 15.2%) and a strong response to thrombin-activating peptide. Leukocyte yield was 93.0% (90.1 to 95.7%) with no activation or cell death. Global removal ratios were 98.3% (97.8 to 98.9%), 98.2% (96.9 to 98.8%), and 88.3% (86.6 to 90.7%) for heparin, albumin, and fibrinogen, respectively. The processing times were 4.4 min (4.2 to 4.6 min) and 4.4 min (4.2 to 4.7 min) for the first and second cycles, respectively. Conclusions This study demonstrated the performance of the SAME device. Platelets and red blood cells were salvaged without significant impact on cell integrity and function. In the meantime, leukocytes were not activated, and the washing quality of the device prevented reinfusion of high concentrations of heparin and plasma proteins. Editor’s Perspective What We Already Know about This Topic What This Article Tells Us That Is New

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Anesthesiology and Pain Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3