Affiliation:
1. Department of Orthopaedics Surgery, The Second Affiliated Hospital and Yuying Children’s Hospital of Wenzhou Medical University, Zhejiang, China
2. West China Hospital of Sichuan University, Chengdu, China
Abstract
Object:
A novel technique, percutaneous elastic stable intramedullary nail fixation (ESIN), proposed by our team for the treatment of anterior pelvic ring injury. Finite element analysis and retrospective case-control study were used to compare biomechanical properties and clinical outcomes between ESIN and other techniques.
Methods:
Four groups of finite element models of pelvic anterior ring injury were simulated, including ESIN (model A), retrograde transpubic screw fixation (RTSF, model B), subcutaneous internal fixator (model C), and external fixator (model D), and a vertical downward load of 500 N was applied to the S1 vertebral endplate. Stress and displacement distributions of intact pelvis, displacement distributions of pubic fracture fragments, and stress distributions of fixation devices were analyzed. Then 31 patients with anterior pelvic ring injury (15 in the ESIN group and 16 in the RTSF group) were reviewed. Clinical outcomes were evaluated at the final follow-up. Postoperative complications were also recorded.
Results:
Under 500N loading, the intact stability of the pelvis was compared as follows: model B (20.58 mm, 121.82 MPa), model A (20.80 mm, 129.97 MPa), model C (22.02 mm, 141.70 MPa), and model D (22.57 mm, 147.06MPa). The regional stability of superior pubic ramus was compared as follows: model B (9.48 mm), model A (10.16 mm), model C (10.52 mm), and model D (10.76 mm). All 31 patients received follow-up at least 12 months post-surgery (range 12–20 months). Age, sex, injury mechanism, fracture type, time between the injury and operation, American Society of Anesthesiologists score, intraoperative blood loss, hospital stay, follow-up period, time to union and Majeed scores did not differ significantly between the two groups (P>0.05). However, the differences in the duration of unilateral surgery, unilateral intraoperative fluoroscopy and one-time success rate were significant (P<0.05).
Conclusions:
With sufficient biomechanical stability and minimally invasive advantage, the percutaneous technique using ESIN can be used to successfully treat anterior pelvic ring injuries. In addition, advantages over RTSF include a shorter duration of surgery, reduced requirement for intraoperative fluoroscopy and a higher one-time success rate. ESIN therefore constitutes a good alternative to RTSF.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献