Abstract
Background
Acoustic reflectometry allows the construction of a one-dimensional image of a cavity, such as the airway or the esophagus. The reflectometric area-distance profile consists of a constant cross-sectional area segment (length of endotracheal tube), followed either by a rapid increase in the area beyond the carina (tracheal intubation) or by an immediate decrease in the area (esophageal intubation).
Methods
Two hundred adult patients were induced and intubated, without restrictions on anesthetic agents or airway adjunct devices. A two-microphone acoustic reflectometer was used to determine whether the breathing tube was placed in the trachea or esophagus. A blinded reflectometer operator, seated a distance away from the patient, interpreted the acoustic area-distance profile alone to decide where the tube was placed. Capnography was used as the gold standard.
Results
Of 200 tracheal intubations confirmed by capnography, the reflectometer operator correctly identified 198 (99% correct tracheal intubation identification rate). In two patients there were false-negative results, patients with a tracheal intubation were interpreted as having an esophageal intubation. A total of 14 esophageal intubations resulted, all correctly identified by reflectometry, for a 100% esophageal intubation identification rate.
Conclusions
Acoustic reflectometry is a rapid, noninvasive method by which to determine whether breathing tube placement is correct (tracheal) or incorrect (esophageal). Reflectometry determination of tube placement may be useful in airway emergencies, particularly in cases where visualization of the glottic area is not possible and capnography may fail, as in patients with cardiac arrest.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Anesthesiology and Pain Medicine
Cited by
28 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献