Sevoflurane Breakdown Produces Flammable Concentrations of Hydrogen

Author:

Dunning Marshall B.1,Bretscher Lynn E.2,Arain Shahbaz R.3,Symkowski Yanina4,Woehlck Harvey J.5

Affiliation:

1. Professor of Medicine and Physiology.

2. Research Scientist, Department of Biophysics.

3. Research Fellow, Department of Anesthesiology.

4. Former Research Support, Department of Anesthesiology.

5. Professor of Anesthesiology, Medical College of Wisconsin.

Abstract

Background Fires, explosions, and extreme heat production may occur when sevoflurane reacts with desiccated barium hydroxide lime. The identity of the flammable gas has not previously been published, although carbon monoxide, methanol, formaldehyde, and methyl formate have been identified in low quantities. Methods The authors reacted sevoflurane with excess desiccated barium hydroxide lime or soda lime at 55 degrees, 100 degrees, 200 degrees, 300 degrees, and 400 degrees C. Formaldehyde, methanol, sodium formate, and hexafluoroisopropanol were reacted with barium hydroxide lime at 300 degrees or 400 degrees C. The authors measured hydrogen production by gas chromatography with a thermal conductivity detector and calculated the molar yield of hydrogen produced. Results Up to 3 moles of hydrogen were produced per mole of sevoflurane degraded. Each mole of formaldehyde produced up to 2 moles of hydrogen at 400 degrees C. Formate and hexafluoroisopropanol produced up to 1 mole of hydrogen each at 400 degrees C. More than 2 moles of hydrogen were produced by methanol at 400 degrees C. Soda lime and barium hydroxide lime produced similar amounts of hydrogen from sevoflurane above 200 degrees C, but barium hydroxide lime produced more than soda lime at lower temperatures. The temperature above which large amounts of hydrogen were produced seemed to be 300 degrees C. Conclusions Up to 3 moles of hydrogen are produced by the chemical reaction of sevoflurane with heated, desiccated absorbent. The high yield, ease of ignition, and low tissue solubility of hydrogen make it the most likely fuel in anesthesia machine fires due to the reaction of sevoflurane with desiccated absorbent.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Anesthesiology and Pain Medicine

Reference15 articles.

Cited by 12 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3