Author:
Iijima Takehiko,Iwao Yasuhide,Sankawa Hiroshi
Abstract
Background
Pulse dye-densitometry (PDD) is a newly developed technique for monitoring the arterial concentration of indocyanine green. Using this method, circulating blood volume (CBV) can be calculated without using radioisotopes. In this study, the CBV value obtained by PDD was validated by comparison using the human serum albumin ((131)I-HSA) dilution method.
Methods
Eleven healthy volunteers underwent placement of cannulae into the radial artery and antecubital vein for withdrawal of blood samples and injection of indicator. Probes for PDD were attached to the right nostril and the right index finger. Indocyanine green (20 mg), dissolved in 4 ml water, and 25 microCi (131)I-HSA in 1 ml distilled water were injected simultaneously into the left antecubital vein. Blood samples were withdrawn 3, 6, 10, 20, 30, and 45 min after injection, then processed for spectrophotometric measurement of indocyanine green and scintillation counting.
Results
The blood dye concentration correlated well with the values obtained by PDD (r=0.986, imprecision 0.04+/-0.11 mg/l, 10.0+/-31%. The imprecision of the CBV value obtained by PDD (nose probe) and by the (131)I-HSA dilution method was 3.99+/-10.54%, 0.259+/-0.593 l. The imprecision of the CBV obtained by in vitro spectrophotometry compared with PDD was 2.47+/-9.00%, 0.100+/-0.446 l.
Conclusions
This newly developed, less invasive method can measure CBV with an imprecision of 3.99+/-10.54%, 0.259+/-0.593 l (nose probe), and thus is also as accurate as the conventional radioisotope method.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Anesthesiology and Pain Medicine
Cited by
83 articles.