Closed-loop Control of Anesthesia Using Bispectral Index

Author:

Absalom Anthony R.1,Sutcliffe Nicholas2,Kenny Gavin N.3

Affiliation:

1. Clinical Lecturer, University Department of Anaesthesia, Glasgow Royal Infirmary. Current position: Consultant Anaesthetist, Department of Anaesthesia, Norfolk and Norwich University Hospital, Norwich, United Kingdom.

2. Consultant Anaesthetist, Health Care International (Scotland) Ltd., Glasgow, United Kingdom.

3. Professor, University Department of Anaesthesia, Glasgow Royal Infirmary.

Abstract

Background The Bispectral Index (BIS) is an electroencephalogram-derived measure of anesthetic depth. A closed-loop anesthesia system was built using BIS as the control variable, a proportional-integral-differential control algorithm, and a propofol target-controlled infusion system as the control actuator. Closed-loop performance was assessed in 10 adult patients. Methods Ten adult patients scheduled to undergo elective hip or knee surgery were enrolled. An epidural cannula was inserted, and 0.5% bupivacaine was used to provide anesthesia to T8 before general anesthesia was induced using the propofol target-controlled infusion system under manual control. After the start of surgery, when anesthesia was clinically adequate, automatic control of anesthesia was commenced using the BIS as the control variable. Adequacy of anesthesia during closed-loop control was assessed clinically and by calculating the median performance error, the median absolute performance error, and the mean offset of the control variable. Results The median performance error and the median absolute performance error were 2.2 and 8.0%, respectively. Mean offset of the BIS from the set point was 0.9. Cardiovascular parameters were stable during closed-loop control. Operating conditions were adequate in all patients but one, who began moving after 45 min of stable anesthesia. No patients reported awareness or recall of intraoperative events. In three patients, there was oscillation of the measured BIS around the set point. Conclusions The system was able to provide clinically adequate anesthesia in 9 of 10 patients. Further studies are required to determine whether control performance can be improved by alterations to the gain factors or by using an effect site-targeted, target-controlled infusion propofol system.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Anesthesiology and Pain Medicine

Reference22 articles.

Cited by 160 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Closed-loop anesthesia: foundations and applications in contemporary perioperative medicine;Journal of Clinical Monitoring and Computing;2024-01-06

2. Controlling Human Physiology via Drugs and Devices;Reference Module in Materials Science and Materials Engineering;2024

3. Closed Loop Anesthesia;2023 International Conference on Intelligent Technologies for Sustainable Electric and Communications Systems (iTech SECOM);2023-12-18

4. Multivariable discrete MPC with exponential cost function on the automation of the anesthesia process;International Journal of Robust and Nonlinear Control;2023-10-03

5. Closed-loop control of anesthetic state in nonhuman primates;PNAS Nexus;2023-09-29