Closed-loop Control of Anesthesia Using Bispectral Index

Author:

Absalom Anthony R.1,Sutcliffe Nicholas2,Kenny Gavin N.3

Affiliation:

1. Clinical Lecturer, University Department of Anaesthesia, Glasgow Royal Infirmary. Current position: Consultant Anaesthetist, Department of Anaesthesia, Norfolk and Norwich University Hospital, Norwich, United Kingdom.

2. Consultant Anaesthetist, Health Care International (Scotland) Ltd., Glasgow, United Kingdom.

3. Professor, University Department of Anaesthesia, Glasgow Royal Infirmary.

Abstract

Background The Bispectral Index (BIS) is an electroencephalogram-derived measure of anesthetic depth. A closed-loop anesthesia system was built using BIS as the control variable, a proportional-integral-differential control algorithm, and a propofol target-controlled infusion system as the control actuator. Closed-loop performance was assessed in 10 adult patients. Methods Ten adult patients scheduled to undergo elective hip or knee surgery were enrolled. An epidural cannula was inserted, and 0.5% bupivacaine was used to provide anesthesia to T8 before general anesthesia was induced using the propofol target-controlled infusion system under manual control. After the start of surgery, when anesthesia was clinically adequate, automatic control of anesthesia was commenced using the BIS as the control variable. Adequacy of anesthesia during closed-loop control was assessed clinically and by calculating the median performance error, the median absolute performance error, and the mean offset of the control variable. Results The median performance error and the median absolute performance error were 2.2 and 8.0%, respectively. Mean offset of the BIS from the set point was 0.9. Cardiovascular parameters were stable during closed-loop control. Operating conditions were adequate in all patients but one, who began moving after 45 min of stable anesthesia. No patients reported awareness or recall of intraoperative events. In three patients, there was oscillation of the measured BIS around the set point. Conclusions The system was able to provide clinically adequate anesthesia in 9 of 10 patients. Further studies are required to determine whether control performance can be improved by alterations to the gain factors or by using an effect site-targeted, target-controlled infusion propofol system.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Anesthesiology and Pain Medicine

Reference22 articles.

Cited by 158 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3