Evaluation of a Cerebral Oximeter as a Monitor of Cerebral Ischemia during Carotid Endarterectomy

Author:

Samra Satwant K.1,Dy Eric A.2,Welch Kathy3,Dorje Pema4,Zelenock Gerald B.5,Stanley James C.5

Affiliation:

1. Professor.

2. Research Assistant.

3. Adjunct Assistant Professor of Biostatistics, Department of Biostatistics.

4. Assistant Professor, Department of Anesthesiology.

5. Professor, Department of Surgery (Vascular Surgery).

Abstract

Background Stroke is an important contributor to perioperative morbidity and mortality associated with carotid endarterectomy (CEA). This investigation was designed to compare the performance of the INVOS-3100 cerebral oximeter to neurologic function, as a means of detecting cerebral ischemia induced by carotid cross-clamping, in patients undergoing carotid endarterectomy with cervical plexus block. Methods Ninety-nine patients undergoing 100 CEAs with regional anesthesia (deep or superficial cervical plexus block) were studied. Bilateral regional cerebrovascular oxygen saturation (rSO2) was monitored using the INVOS-3100 cerebral oximeter. Patients were retrospectively assigned to one of two groups: those in whom a change in mental status or contralateral motor deficit was noted after internal carotid clamping (neurologic symptoms; n = 10) and those who did not show any neurologic change (no neurologic symptoms; n = 90). Data from 94 operations (neurologic symptoms = 10 and no neurologic symptoms = 84) were adequate for statistical analyses for group comparisons. A relative decrease in ipsilateral rSO2 after carotid occlusion (calculated as a percentage of preocclusion value) during all operations (n = 100) was also calculated to determine the critical level of rSO2 decrease associated with a change in neurologic function. Results The mean (+/- SD) decrease in rSO2 after carotid occlusion in the neurologic symptoms group (from 63.2 +/- 8.4% to 51.0 +/- 11.6%) was significantly greater (P = 0.0002) than in the no neurologic symptoms group (from 65.8 +/- 8.5% to 61.0 +/- 9.3%). Logistic regression analysis used to determine if a change in rSO2, calculated as a percentage of preclamp value, could be used to predict change in neurologic function was highly significant (likelihood ratio chi-square = 13.7; P = 0.0002). A 20% decrease in rSO2 reading from the preclamp baseline, as a predictor of neurologic compromise, resulted in a sensitivity of 80% and specificity of 82.2%. The false-positive rate using this cutoff point was 66.7%, and the false-negative rate was 2.6%, providing a positive predictive value of 33.3% and a negative predictive value of 97.4%. Conclusion Monitoring rSO2 with INVOS-3100 to detect cerebral ischemia during CEA has a high negative predictive value, but the positive predictive value is low.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Anesthesiology and Pain Medicine

Reference25 articles.

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3