The ProSeal Laryngeal Mask Airway

Author:

Brimacombe Joseph1,Keller Christian2

Affiliation:

1. Clinical Professor, University of Queensland, Department of Anaesthesia and Intensive Care, Cairns Base Hospital, Cairns, Australia.

2. Consultant Anaesthetist, Department of Anaesthesia and Intensive Care Medicine, Leopold-Franzens University, Innsbruck, Austria.

Abstract

Background The ProSeal laryngeal mask airway (PLMA) is a new laryngeal mask device with a modified cuff to improve seal and a drainage tube to provide a channel for regurgitated fluid and gastric tube placement. In the present randomized, crossover study, the authors tested the hypothesis that ease of insertion, airway sealing pressure, and fiberoptic position differ between the PLMA and the standard laryngeal mask airway (LMA). For the PLMA, we also assess ease of gastric tube placement and the efficacy of an introducer tool. Methods Sixty paralyzed, anesthetized adult patients were studied. Both devices (only size 4) were inserted into each patient in random order. Airway sealing pressure and fiberoptic position were determined during cuff inflation from 0 to 40 ml in 10-ml increments. Gastric tube insertion was attempted with the PLMA if there was no gas leak from the drainage tube. In 60 additional patients, ease of insertion for the PLMA was compared with and without an introducer. Results First-time success rates were higher (60 of 60 vs. 52 of 60; P = 0.003) and the effective airway time shorter (9 +/- 3 s vs20 +/- 18 s; P < 0.0001) for the LMA. There were no failed uses of either device within three attempts. Airway sealing pressure was 8-11 cm H2O higher for the PLMA at all cuff volumes (P < 0.00001) and was higher in females for both devices. Fiberoptic position was better with the LMA at all cuff volumes (P < 0.00001), but vocal cord visibility was similar (LMA, 59 of 60; PLMA, 56 of 60). For the PLMA, gastric tube placement was successful in 58 of 58 patients and took 9 +/- 5 s. First-time success rates were higher (59 of 60 vs53/60; P = 0.03) and the effective airway time shorter (15 +/- 13 s vs 23 +/- 18 s; P = 0.008) with the introducer. Conclusion The PLMA is capable of achieving a more effective seal than the LMA and facilitates gastric tube placement, but it is more difficult to insert unless an introducer tool is used. When correctly positioned, the PLMA isolates the glottis from the upper esophagus with possible implications for airway protection.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Anesthesiology and Pain Medicine

Reference11 articles.

Cited by 241 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3