Affiliation:
1. Associate Professor.
2. Assistant Professor.
3. Professor and Chairman, Department of Statistics, and Professor, Department of Anesthesia and Critical Care.
4. Professor and Vice Chair, Department of Anesthesia and Critical Care.
Abstract
Background
Unexpected awareness is a rare but well-described complication of general anesthesia that has received increased scientific and media attention in the past few years. Transformed electroencephalogram monitors, such as the Bispectral Index monitor, have been advocated as tools to prevent unexpected recall.
Methods
The authors conducted a power analysis to estimate how many patients would be needed in an appropriately powered study to demonstrate the Bispectral Index monitor reduces awareness, as well as a cost analysis to assess the cost of using the monitor for this purpose alone.
Results
If unexpected recall is rare (1 in 20,000), it will require a large study to demonstrate that the monitor reduces awareness (200,000-800,000 patients), and the cost of using it for this purpose alone would be high ($400,000 per case prevented). If awareness is common (1 in 100), then the number of patients needed in a study to demonstrate that the monitor works becomes tractable (1,000-4,000 patients), and the cost of using the monitor for this purpose alone becomes lower ($2,000 per case prevented). Because there are reported cases of awareness despite Bispectral Index monitoring, the authors are certain that the effectiveness of the monitor is less than 100%. As the performance of the monitor decreases from 100%, the size of the study needed to demonstrate that it works increases, as does the cost of using it to prevent awareness.
Conclusion
The contention that Bispectral Index monitoring reduces the risk of awareness is unproven, and the cost of using it for this indication is currently unknown.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Anesthesiology and Pain Medicine
Cited by
91 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献