Author:
Reali-Forster Chiara,Kolobow Theodor,Giacomini Matteo,Hayashi Tomayoshi,Horiba Koji,Ferrans Victor J.
Abstract
Background
A new endotracheal tube (ETT) was fabricated and tested in sheep. It had no tracheal cuff; airway seal was achieved at the level of the glottis through a no-pressure seal made of "gills"; the laryngeal portion was oval-shaped; and the wall thickness was reduced to 0.2 mm.
Methods
Sheep were tracheally intubated either with a standard tube or with the new tube, and their lungs were mechanically ventilated for 1 or 3 days. Air leak was recorded at different peak inspiratory pressures (PIPs). Liquid seepage into the trachea was assessed using an indicator dye. Tracheolaryngeal lesions were scored grossly and histologically.
Results
There was no air leak up to 40 cmH2O of PIP, in either group, in short- and long-term studies. Methylene blue leaked across the cuff in two sheep with standard ETTs. No dye leaked across the gills with the new ETTs. In the new ETT group, the trachea appeared better preserved, grossly and histologically, than in the standard ETT group at both 1 and 3 days (P < 0.05). At day 1, the larynx and vocal cords appeared grossly less injured in the new ETT group (P < 0.05), whereas there was no difference at day 3. Histology did not show significant difference on vocal cords, epiglottis, and larynx between the two groups at any time.
Conclusions
The novel, no-pressure seal design of the new ETT is highly effective in preventing air leak and aspiration. It causes no significant tracheal injury.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Anesthesiology and Pain Medicine
Cited by
47 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献