1. Department of Nuclear Medicine, Cerrahpasa Medical Faculty, Istanbul University-Cerrahpasa, Fatih, Istanbul, Turkey
Abstract
Objective
The aim of this study was to perform and evaluate PET/computed tomography acceptance tests separately using American Association of Physicists in Medicine (AAPM) Task Group 126 and National Electrical Manufacturers Association (NEMA) methods.
Materials and methods
Measurements of sensitivity, spatial resolution, count rate performance and scatter fraction, the accuracy of corrections for count losses and randoms, and image quality were obtained according to NEMA NU-2018. Likewise, the performance tests were made using the AAPM Task Group 126 method, and the results were compared with NEMA NU-2018.
Results
The sensitivity at the isocenter was 8.87 cps/kBq according to NEMA and 7.60 cps/kBq by using the AAPM Task Group 126. For the spatial resolution, the full width at half maximum (FWHM) and FWTM values were 4.34 mm and 6.78 mm at 1 cm radial offset by NEMA, while AAPM Task Group 126 yielded FWHM and FWTM values of 4.42 mm and 8.14 mm, respectively. In the image quality, NEMA exhibited hot lesions contrast of 40.8, 56.7, 69.9, and 77.3 for 10, 13, 17, and 22 mm spheres, respectively. As a ratio to 25 mm, the “Hot” max standard uptake values by AAPM Task Group 126 were found to be 1, 1.1, 1.37, and 1.68 for 8, 12, 16, and 25 mm lesions, respectively.
Conclusion
Acceptance tests using NEMA are of high relevance and convenience for the reliability of the results. Alternatively, AAPM Task Group 126 seems convenient and more economical to apply with reliable outcomes for the equivalent tests.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Radiology, Nuclear Medicine and imaging,General Medicine
Reference9 articles.
1. PET/CT acceptance testing and quality assurance.;Mawlawi;AAPM Report No. 126,2019
2. A method for measuring the absolute sensitivity of positron emission tomographic scanners.;Bailey;Eur J Nucl Med Mol Imaging,1991
3. Discovery molecular imaging digital ready PET/CT performance evaluation according to the NEMA NU2-2012 Standard.;Michopoulou;Nucl Med Commun,2019
4. Comparison of NEMA characterizations for Discovery MI and Discovery MI-DR TOF PET/CT systems at different sites and with other commercial PET/CT systems.;Chicheportiche;EJNMMI Physics,2020
5. PET performance measurements for an LSO-based combined PET/CT scanner using the National Electrical Manufacturers Association NU 2-2001 Standard.;Erdi;J Nucl Med,2004
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献