What Are the Ready-to-Use Endoscope Channels Hiding?

Author:

Garcia Naiara BussolottiORCID,Oliveira Adriana Cristina de

Abstract

Contamination due to failures or omissions in the reprocessing steps of gastrointestinal endoscopes is common in clinical practice. Ensuring the proper execution of each step is a challenge for reprocessing personnel. This cross-sectional study was conducted in an endoscopy setting between March and May 2021. We performed interviews about reprocessing practices, analyzed the life history of the equipment, and performed inspections through a borescope video of gastrointestinal endoscope channels that were stored and ready for use. A borescope is a complementary tool used to validate endoscope reprocessing, evaluate the internal visualization of channels, and identify changes that can compromise the safety of its use, which are often not detected in the leak test. Thirteen biopsy channels from stored gastrointestinal endoscopes were inspected. We found that 85% had stains and grooves, 69% contained moisture, and 46% had debris. There was at least one noncompliance issue in all of the channels inspected.