Safety of Magnetic Resonance Imaging in Orthopaedic Trauma Patients With External Fixation: A Two-Center Case Series

Author:

Borgida Jacob S.12ORCID,Wagner Robert Kaspar12,Wong Alice W.12,Yee Seonghwan3ORCID,Husseini Jad4ORCID,Aneja Arun2ORCID,Harris Mitchel B.2ORCID,Ly Thuan V.12ORCID

Affiliation:

1. Harvard Orthopaedic Trauma Initiative, Massachusetts General Hospital, Boston, MA;

2. Department of Orthopaedic Surgery, Massachusetts General Hospital, Harvard Medical School, Boston, MA;

3. Department of Radiology, Massachusetts General Hospital, Boston, MA; and

4. Division of Musculoskeletal Imaging and Intervention, Department of Radiology, Massachusetts General Hospital, Boston, MA.

Abstract

OBJECTIVES: To report on adverse events during magnetic resonance imaging (MRI) in patients with external fixators. METHODS: Design: Retrospective case series. Setting: Two Level 1 trauma centers. Patient Selection Criteria: Patients with external fixators on the appendicular skeleton or pelvis undergoing MRI between January 2005 and September 2023. Outcome Measures and Comparisons: Adverse events, defined as any undesirable event associated with the external fixator being inside or outside the MRI bore during imaging, including (subjective) heating, displacement or pullout of the external fixator, or early MRI termination for any reason. RESULTS: A total of 97 patients with 110 external fixators underwent at least one MRI scan with an external fixator inside or outside of the MRI bore. The median age was 51 years (interquartile range: 39–63) and 56 (58%) were male. The most common external fixator locations were the ankle (24%), knee (21%), femur (21%), and pelvis (19%). The median duration of the MRI was 40 minutes (interquartile range: 26–58), 86% was performed using 1.5-Tesla MRI, and 14% was performed using 3.0-Tesla MRI. Ninety-five percent of MRI was performed for the cervical spine/head. Two MRI scans (1.6%), one of the shoulder and one of the head and cervical spine, with the external fixator outside of the bore were terminated early because of patient discomfort. There were no documented events of displacement or pullout of the external fixator. CONCLUSIONS: These findings suggest that MRI scans of the (cervical) spine and head can be safely obtained in patients with external fixators on the appendicular skeleton or pelvis. Given the low numbers of MRI scans performed with the external fixator inside the bore, additional studies are necessitated to determine the safety of this procedure. The results from this study can aid orthopaedic surgeons, radiologists, and other stakeholders in developing local institutional guidelines on MRI scanning with external fixators in situ. LEVEL OF EVIDENCE: Prognostic Level III. See Instructions for Authors for a complete description of levels of evidence.

Publisher

Ovid Technologies (Wolters Kluwer Health)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3