Assessment of coagulation assays on Roche Cobas t711 analyzer: performance and clinical implications

Author:

Ordóñez-Robles María1,Pons-Belda Oscar D.1,Moína María José1,Bernardo-Gutiérrez Ángel12,Prieto-García Belén1

Affiliation:

1. Laboratorio de Medicina

2. Servicio de Hematología, Hospital Universitario Central de Asturias (HUCA), Spain

Abstract

Objectives We performed an analytical assessment of five coagulation tests [i.e. prothrombin time (PT), activated partial thromboplastin time (aPTT), fibrinogen, thrombin time (TT) and D-dimer] on the Roche Cobas t711 analyzer and a comparison study with the methodology in use at our laboratory (i.e. Werfen ACL Top 750 analyzer), expanding the analysis to the clinical implications of Cobas t711 implementation. Methods Imprecision studies were performed following the Clinical and Laboratory Standards Institute (CLSI) H57 A:2008 guideline. Linearity of D-dimer and fibrinogen tests was analysed according to the CLSI EP06-A: 2003 recommendations. For method comparison, the results were analyzed using the Bland-Altman plot and Passing-Bablok regression. Results Imprecision met manufacturer claims for PT, aPTT and TT. D-dimer and fibrinogen tests showed a coefficient of variation (CV)% over manufacturer claims at certain concentration levels. Linearity ranges could not be verified. Comparison study revealed that results are not interchangeable for any test, a lower correlation for aPTT test and lower D-dimer results from Roche Cobas t711. Conclusion The strength of this study relies on the analysis of the clinical implications of reporting Cobas t711 results compared to those obtained with the methodology in use at our laboratory. Different sensibility to factor deficiency, anticoagulant therapy and interferences might explain lower correlation rates obtained for the aPTT test. Different monoclonal antibodies used for D-dimer determination might explain the lower results obtained with the Cobas t711 analyzer. This aspect needs further studies given the relevance of D-dimer test to exclude thrombotic events and reinforces the need of harmonization in the haemostasis laboratory.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Hematology,General Medicine

Reference23 articles.

1. Laboratory assessment of the anticoagulant activity of direct oral anticoagulants: a systematic review;Samuelson;Chest,2017

2. Monitoring of unfractionated heparin in critically ill patients;Aarab;Neth J Med,2013

3. Laboratory hemostasis: from biology to the bench;Lippi;Clin Chem Lab Med,2018

4. Reflections on the next generation of hemostasis instrumentation. A glimpse into the future?;Lippi;J Lab Med,2016

5. Verification of examination procedures in clinical laboratory for imprecision, trueness and diagnostic accuracy according to ISO 15189:2012: a pragmatic approach;Antonelli;Clin Chem Lab Med,2017